Specification

Specification

Product SpecificationsValue
Printing Speed 1,2,3cm/min,( Real time)
30 cm/min (Trace, 2,3 cm/min setting)
20 cm/min (Trace, 1 cm/min setting)
Fetal Heart Rate Input Signal : Ultrasound pulsed doppler
FHR detection method : Autocorrelation
FHR range : 50~210
FHR accuracy : 120~160bpm : ±1 bpm
Except 120~160 : ±2 bpm
Ultrasound Tranducer Operating mode : PWD Mode
Transducer Type : 9-crystal
Ultrasound frequency : 1.0MHz
Pulse Repetition Frequency : 3125Hz
Spatial-Peak Temporal Average Intensity : <10mW/cm2
Display 7” Wide (800 X 480)
User Interface Short Key
Printer Resolution Method : Thermal Array Print
Type: Z-folder
"Print speed : 1,2,3cm/min,( Real time)
30 cm/min (Trace, 2,3 cm/min setting)
20 cm/min (Trace, 1 cm/min setting)"
Paper feeding function
Auto CTG Analysis Average Baseline FHR
Number of TOCO
Number of Acceleration
Number of Deceleration : Late, Early, Variable
High/Low Episode
Short term Variability
Signal Loss
* CTG Analysis results is printed out every 10 minutes
Uterine Contraction Input Source : External Transducer
Reference Control : One touch switch/ Auto zeroing
Measurement range : 0~99
Line Power DC Input: 18V, 2.8A/ Adaptor “USE THE ONLY BridgePower Corp BPM050S18F02”
Battery type Lithium ion
Battery Capacity Li-ion : 4 hours ( charging)
2 hours(discharging)
Data Storage Storage for 72 hours
Environmental Operation • Operating Temperature: 10°C to 40°C (50°F to 104°F)
• Storage Temperature: - 10°C to 60°C (14°F to 140°F)
• Operating Humidity: 30% to 85% RH
• Storage Humidity: 20% to 95% RH
• Operating Attitude: 70(700) to 106Kpa(1060mbar)
• Storage Attitude: 50(500) to 106Kpa(1060mbar)
Dimensions 296(w)x.305.5(H)x97.5(D)(Approx. 2.9Kg)
Default Alarm Settings Alarm Parameter: ON
Alarm Limit: 160, 120
Alarm Level: Medium